Culture Myograph System Model 202CM
2.2.1 Setting up the Complete Culture Myograph System 202CM
This section describes how to connect the cables in the culture myograph system
as illustrated in Fig. 2.1. Before connecting any of the myograph equipment, ensure
that the frame grabber card is installed into the PCI slot of the computer. Please see
installation instructions in Section 2.2.2.
1. Ensure that the heat controller is switched off, on the rear panel before proceeding
with the connection procedure.
2. Connect the loose cables from the four culture myograph units to the front panel of
the heat controller.
3. Connect the power cord to the power inlet on the back panel of the heat controller.
4. Connect the Heat Controller to the PC with the loose USB cable using the USB port
on the Heat Controller and a free USB port on the PC.
5. Connect the DMT Microscope to a USB port on the computer using the xed USB
cord.
6. Turn on the power switch of the heat controller.
7. The Power ON LED on the front panel of the heat controller should be lit indicating
that the heat controller is on.
6. Connect the DSUB 9-pol cable from the DMT microscope to the 9-pol port on the
frame grabber card installed in the computer.
7. Turn on the computer.
2.2.2 Installation of Frame Grabber, Drivers and VediView
™
This section describes how to install the Euresys Picolo frame grabber PCI card in
your computer along with drivers and VediView
™
software.
Note:
If you have purchased a computer from DMT in conjunction with your Culture
Myograph 202CM System the card, drivers and VediView
™
software have already
been installed for you.
1. Turn off your computer and install the Euresys Picolo frame grabber PCI card as
described in the computer or motherboard manual.
2. Turn on the culture myograph system and the computer. Typically the Windows
software automatically detects the installed frame grabber and starts up a “Guide
to install new hardware”. Skip the automatic installation (Click “Cancel”) and
proceed.